Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix A: Using switch matrices
4200A-901-01 Rev. C / February 2017
A-31
Set up and run a switch matrix in Clarius
To set up and run the
connectpins
action:
1. Choose
Select
.
2. Select
Actions
.
3. Search for
connectpins
.
4. Select the
connectpins
action.
5. Select
Add
.
6. In the project tree, select the
connectpins
action.
7. Select
Configure
. The parameter settings are displayed, as shown below.
8. Set Pin2 to
6
. This connects SMU2 to point 6.
9. Select
OpenAll
to open all matrix card switches.
10. Set Pin5 to
3
. This connects GNDU to pin 5.
11. Leave all other pin settings at 0 to indicate that no connection will be made.
Figure 545: connectpins settings
12. Select
Run
. The 4200A-SCS connects to the DUT as shown in
(on page A-