Appendix B: Using a Model 590 C-V Analyzer
Model 4200A-SCS Parameter Analyzer Reference Manual
B-30
4200A-901-01 Rev. C / February 2017
If the file defined for
CabCompFile
does not exist, or there is no path specified (null string), the
default compensation parameters are used. When entering the path, use two backslash (
\\
)
characters to separate each directory. For example, if your cable file is in
C:\calfiles\590cal.dat
, you enter the following:
C:\\calfiles\\590cal.dat
If a switch matrix to route signals is being controlled by a connection action (for example,
connect
),
there is no need to connect
InputPin
and
OutPin
. Set these parameters to
0
.
For
Range
, the measurement range values are shown in the following table.
Range
100 kHz
1 MHz
1
2 pF / 2
μs
20 pF / 200 μs
2
20 pF / 20 μs
20 pF / 200 μs
3
200 pF / 200 μs
200 pF / 2 ms
4
2 nF / 2 ms
2 nF / 20 ms
The reading rates and resolutions for the
ReadingRate
parameter are described in the following
table.
Reading rate
Nominal reading rate
(per second)
Display readings
Resolution
(digits)
0
1000
C
3.5
1
75
C,G,V
3.5
2
18
C,G,V
4.5
3
10
C,G,V
4.5
4
1
C,G,V
4.5
Returned values are placed in the Analyze sheet:
•
0
: OK.
•
-10000 (INVAL_INST_ID)
: The specified instrument ID does not exist.
•
-10020 (COMP_FILE_ACCESS_ERR)
: There was an error accessing the specified cable
compensation file.
•
-10021 (COMP_FILE_NOT_EXIST)
: The specified compensation file does not exist.
•
-10022 (KI590_NOT_IN_KCON)
: There is no CMTR defined in your system configuration.
•
-10023 (KI590_MEAS_ERROR)
: A measurement error occurred.
•
-10090 (GPIB_ERROR_OCCURRED)
: A GPIB communications error occurred.
•
-10091 (GPIB_TIMEOUT)
: A timeout occurred during communications.
•
-10101 (ARRAY_SIZE_TOO_SMALL)
: The specified value for
Csize
,
G_or_Rsize
,
Vsize
, or
Tsize
was too small for the number of steps in the sweep.
•
-10102 (ERROR_PARSING)
: There was an error parsing the response from the 590.
•
-10104 (USER_CANCEL)
: The user canceled the correction procedure.