Model 4200A-SCS Parameter Analyzer Reference Manual
Section 4: Multi-frequency capacitance-voltage unit
4200A-901-01 Rev. C / February 2017
4-21
4210-CVU project example
The following example shows you how to use the 4210-CVU in a project to make a capacitance
measurement.
The example assumes you have a 4210-CVU connected to a test fixture as described in
4210-CVU test connections to a DUT
(on page 4-6).
In this test, you sweep the DC bias from
−
5 V to 5 V in 0.2 V steps, with a 1 MHz capacitance
measurement made at each step.
Select a project
Select a project:
1. Start
Clarius
.
2. Do
(on page 4-10) to correct for offset and gain errors caused by the
connections.
3. Choose
Select
.
4. Select
Projects
to display the Project Library in the center pane.
5. Search for
demo
. The Demo Project is displayed, as shown in the next figure.
Figure 100: Demo Project selected