Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix D: Using a Model 82 C-V System
4200A-901-01 Rev. C / February 2017
D-13
To open and execute the QTsweep UTM:
1. Choose
Select
.
2. In the Test Library, select Custom Test,
Choose a test from the pre-programmed library
(UTM)
.
3. Drag the test to the project tree.
4. Select
Configure
.
5. In the Test Settings pane, under User Libraries, select
ki82ulib
.
6. Under User Modules, select
QTsweep82
.
7. Modify the test parameters as needed. Refer to
(on page D-38) for
parameter definitions.
If you use the parameters shown in the figure below, the Model 82 makes 20 quasistatic
capacitance measurements using 20 mV pulses (
V_Step
) ranging from 0.07 seconds to
1 second (
Delay_Max
).
8. Click
Run
to execute the test.
Figure 590: QTsweep82 user module