Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix D: Using a Model 82 C-V System
4200A-901-01 Rev. C / February 2017
D-35
The analog filter, enabled with
Filter590
, can minimize the amount of noise that appears in the
readings. It does, however, increase the measurement time.
Reading_rate
valid inputs
Reading
rate
Nominal reading rate
(per second)
Readings
Display resolution
(digits)
1
75
C,G,V
3.5
2
18
C,G,V
4.5
3
10
C,G,V
4.5
4
1
C,G,V
4.5
Range590
valid range values
Range
100 kHz
1 MHz
1
2 pF / 2 µs
20 pF / 200 µs
2
20 pF / 20 µs
20 pF / 200 µs
3
200 pF / 200 µs
200 pF / 2 ms
4
2 nF / 2 ms
2 nF / 20 ms
If
CabCompFile
does not exist, or if there is no path specified (null string), the default compensation
parameters are used. When entering the path, be sure to use two
\
characters to separate each
directory. For example, if your cable file is in:
C:\calfiles\82cal.dat
You would enter:
C:\\calfiles\\82cal.dat
If you are controlling a switch matrix to route signals using a connection UTM (for example,
“connect”), you do not need connect
InputPin
and
OutputPin
. Set these parameters to
0
.
The return values from
status
can be:
•
0
: OK.
•
-10000 (INVAL_INST_ID)
: The specified instrument ID does not exist
(INVAL_INST_ID)
:
The specified instrument ID does not exist
•
-10020 (COMP_FILE_ACCESS_ERR)
: There was an error accessing the specified cable
compensation file
•
-10021 (COMP_FILE_NOT_EXIST)
: The specified compensation file does not exist
•
-10045
(
KI82_NOT_IN_KCON
): There is no CMTR defined in your system configuration
•
-10023 (KI590_MEAS_ERROR)
: A measurement error occurred
•
-10101 (ARRAY_SIZE_TOO_SMALL)
: The specified value for
CHF_ArrSize
,
G_or_R_ArrSize
, or
Time_ArrSize
was too small for the number of steps in the sweep
•
-10102 (ERROR_PARSING)
: There was an error parsing the response of the 590.
•
-10104 (USER_CANCEL)
: The user canceled the correction procedure.
•
-10090 (GPIB_ERROR_OCCURRED)
: A GPIB communications error occurred
•
-10100 (INVAL_PARAM)
: An invalid input parameter is specified