Appendix D: Using a Model 82 C-V System
Model 4200A-SCS Parameter Analyzer Reference Manual
D-46
4200A-901-01 Rev. C / February 2017
The following table lists the valid range values for
Range590
.
Range590
valid range values
Range
100 kHz
1 MHz
1
2 pF / 2 µs
20 pF / 200 µs
2
20 pF / 20 µs
20 pF / 200 µs
3
200 pF / 200 µs 200 pF / 2 ms
4
2 nF / 2 ms
2 nF / 20 ms
If
CabCompFile
does not exist, or if there is no path specified (null string), the default compensation
parameters are used. When entering the path, be sure to use two
\
characters to separate each
directory. For example, if your cable file is in:
C:\calfiles\590cal.dat
You would enter:
C:\\calfiles\\590cal.dat
The return values from
status
can be:
•
0
: OK.
•
-10000 (INVAL_INST_ID)
: The specified instrument ID does not exist
•
-10020 (COMP_FILE_ACCESS_ERR)
: There was an error accessing the specified cable
compensation file
•
-10021 (COMP_FILE_NOT_EXIST)
: The specified compensation file does not exist
•
-10023
(
KI590_MEAS_ERROR
): A measurement error occurred
•
-10090 (GPIB_ERROR_OCCURRED)
: A GPIB communications error occurred
•
-10091 (GPIB_TIMEOUT)
: A timeout occurred during communications
•
-10100 (INVAL_PARAM)
: An invalid input parameter is specified
•
-10101
(
ARRAY_SIZE_TOO_SMALL
): The specified value for
CHF_ArrSize
,
G_or_R_ArrSize
,
V_ArrSize, CQS_ArrSize
, or
QT_ArrSize
was too small for the number
of steps in the sweep
•
-10102 (ERROR_PARSING
): There was an error parsing the 590 response
•
-10104 (USER_CANCEL
): The user canceled the correction procedure
•
-10045 (KI82_NOT_IN_KCON
): KI82 is not in KCon
Procedure
1. If set, you are prompted to open the circuit so that an offset capacitance measurement can be
made.
2. If a cable compensation file is specified, the compensation information in that file for the selected
range and frequency will be loaded. If not, instrument default compensation is used.
3. A simultaneous C-V sweep is made.
Also see
None