Appendix F: Set up a probe station
Model 4200A-SCS Parameter Analyzer Reference Manual
F-18
4200A-901-01 Rev. C / February 2017
7. The action
connect
connects the SMUs to the probes for the npn transistor as shown in the
following figure.
Figure 632: Connect SMUs to NPN transistor
8. The test runs
vce-ic-1x
, which generates a collector family of curves (I
C
vs. V
C
) for the
transistor.
9. The action
prober-ss-move
moves the prober to the next subsite.
10. The tests continue with
subsite2
and
subsite3
.
11. After all the subsites have run, the action
prober-separate
separates the prober pins from the
wafer.
12. The action
prober-prompt
displays the message “Wafer Test Complete” at the end of the test.