Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix G: Using a Cascade Microtech PA200 Prober
4200A-901-01 Rev. C / February 2017
G-21
2. From the ProberBench NT window, select
WaferMap
file.
Figure 662: ProberBench NT window
3. From the WaferMap window, from the Mark Dies menu, select
Mark to Skip
.
Figure 663: Mark Dies menu
4. Use Mark to Skip and Mark to Probe to set dies. Click a die in the WaferMap window to either set
or clear the die. The color of the die indicates status (either probe or skip). With Mark to Probe
selected, drag to select multiple dies. With Mark to Skip selected, drag to clear multiple dies.
When done, clear
Mark to Skip
or
Mark to Probe
. Otherwise, the Chuck menu remains grayed.