Appendix G: Using a Cascade Microtech PA200 Prober
Model 4200A-SCS Parameter Analyzer Reference Manual
G-26
4200A-901-01 Rev. C / February 2017
5. Select the
Karl Suss PA200 Prober
as the model.
6. Ensure that the
Number of Pins / Positioners
is correct. The number of pins defined here
determines the pins that are available to assign to a switch matrix card column.
7. Select
Save
.
8. Exit KCon.
Running projects
On the ProberBench NT computer:
1. After the wafer is set up and alignment is complete, select the
File > Project > Save
.
2. Select
File > Map > Save
.
3. Select
File > Table > Save
. The wafer is ready to probe.
4. Place the prober in Run mode.
5. Ensure that the "E" in the
WaferMap
toolbar is selected.
Figure 672: WaferMap toolbar
Clarius
Use Clarius to load and run the
probesites
or
probesubsites
project using the new KCon
configuration file, which allows you to execute the project for this prober.
On the 4200A-SCS:
1. Open Clarius.
2. Choose
Select
.
3. Select
Projects
.
4. Search for
probesubsites
.
5. Drag the
probesubsites
project to the project tree.