Appendix H: Using a Micromanipulator 8860 Prober
Model 4200A-SCS Parameter Analyzer Reference Manual
H-4
4200A-901-01 Rev. C / February 2017
5. Double-click the
pcBridge
icon on the desktop to open the main pcBridge window.
Figure 675: pcBridge icon
Figure 676: Prober setup: Main pcBridge window
6. Select the
Setup
menu. The pcBridge Communications Setup window is displayed.
Figure 677: pcBridge window
7. Use the pcBridge Communications Setup to configure the communications settings. These
settings should be 8 data, 1 stop, no parity, xon/xoff.
•
Interface Type: RS232
•
Baud: 9600
•
Port: COM2
•
Term: cr and lf (termination character of carriage-return and line-feed)
8. Click
OK
.
9. Click the
pcLaunch
icon to open the main pcLaunch window.
Figure 678: pcLaunch icon