Section 4: Multi-frequency capacitance-voltage unit
Model 4200A-SCS Parameter Analyzer Reference Manual
4-82
4200A-901-01 Rev. C / February 2017
mosfet formulas and constants
Formulas and user-defined constants that are used for the application tests are summarized (in
alphabetical order) in the next topics. The formulas and constants are set from the Formulator
settings for the tests.
Formula: AR
Formula name: AR (
aR
)
Units: None
Description: Intermediate parameter for calculation of corrected capacitance.
Formulator entry:
AR = GP_GB-(GP_GB^2 + (2*PI*F_GB*CP_GB)^2)*RS
Simplified equation:
Formula: CADJ
Formula name: CADJ (
CADJ
)
Units: F
Description: Corrected capacitance by compensating series resistance.
Formulator entry:
CADJ = ((GP_GB^2 + (2*PI*F_GB*CP_GB)^2)*CP_GB)/(AR^2 +(2*PI*F_GB*CP_GB)^2)
Simplified equation:
Formula: CFB_CALC
Formula name: CFB_CALC (
CFB
)
Units: F
Description: Flatband capacitance.
Formulator entry:
CFB_CALC = (COX_CALC*ES*AREA/(DEBYEM*1E2))/(C(ES*AREA/(DEBYEM*1E2)))
Simplified equation: