Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-71
Base
The voltage offset (from 0 V) that is the reference for the pulse amplitude.
Amplitude
The amplitude voltage.
Dual Sweep (pulse)
When you select Dual Sweep, the instrument sweeps from start to stop, then from stop to start. When
you clear Dual Sweep, the instrument sweeps from start to stop only.
The amplitude, base, and DC sweeps have an option for Dual Sweep. The figure below shows the
difference between a single pulse sweep and a dual pulse sweep. This figure illustrates a linear
voltage sweep; 0 V to 4 V in 1 V steps.
Figure 248: Single and dual pulse amplitude sweep examples
Force Range (PMU)
The range that is used when sourcing. Select one of the listed ranges. The source remains on the
range that is set. If you are sweeping and a sweep point exceeds the source range capability, the
source outputs the maximum level for that range. This range must be equal to or greater than the
largest value in the sweep.
Note that the 40 V range has a higher maximum output voltage and current, but does not have the
faster transition times of the 10 V range or the lower current measure ranges of the 4225-RPM.