Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-107
Create a custom ITM
You can create a custom interactive test module (ITM) in Clarius. You do not need to create any
external files (such as user modules) to create a custom ITM.
When you create a blank ITM, the number of terminals in the new test are determined by the type of
device the test is placed under.
To create an ITM custom test:
1. Choose
Select
.
2. Highlight a device in the project tree or add a device.
3. If you need to add a device, open the Devices tab and select a device.
4. Select the
Tests
tab.
5. In the Test Library, select
Custom Test
.
6. Select
Add a blank test that can be configured into a DC, Pulse, or CV test (ITM)
.
7. Drag
Custom Test
to the project tree. The test has a red triangle next to it to indicate that it is not
configured.
8. Select
Rename
.
9. Type a name for the test and press
Enter
.
10. Select
Configure
to set up the test.
11. Select the instrument.
12. Refer to
Test and terminal setting descriptions
(on page 6-22) to configure the other options.
13. For each device terminal, ensure that the physical device connections match the device
connections defined in Clarius. If necessary, shut down the instrumentation and correct the
physical connections.
Physical device-terminal connections must accurately match virtual connections to avoid inaccurate
test results and potential device damage.