Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-176
4200A-901-01 Rev. C / February 2017
Export output values to Analyze sheet
For subsite cycling, you can export output values from tests into the subsite Analyze sheet. Each time
a subsite is cycled, the measurements for the output values are placed in the subsite Analyze sheet.
If, for example, the subsite is cycled five times, there will be five measured readings for each output
value.
To select the values to be exported:
1. Select the test in the project tree.
2. Select
Configure
.
3. In the right pane, select the
Test Settings
tab.
4. Select
Output Values
.
5. Select the values to export into the subsite Analyze sheet.
6. Select
OK
.
Run a complex test
This section describes how to run individual tests, devices, subsites, and sites. It also describes how
to run the stress modes.
If Clarius detects an above-normal temperature condition at any SMU, it protects system outputs by
preventing or aborting a test run and reporting the condition in the message area of the Clarius
window. If the condition occurs when a test is attempted, Clarius prohibits execution. If the condition
occurs during a test, Clarius aborts the test.
Run projects, subsites, devices, and tests
You can execute an entire project or individual parts of the project.
While a test is running, you can view test data in the Analyze pane.
The Message area at the bottom of the center pane of Clarius displays the start time, stop time, and
total execution time of the components that were run.
Run projects
Executing an entire project runs its components, including sites, subsites, devices, tests, and actions,
in the order in which they appear in the project tree.
Clarius only runs items that are selected in the project tree and are at a lower level than the
highlighted item.
When you run a project, the data from each test is inserted into its own Analyze Run worksheet. Each
new run creates a new Run History, with it own worksheet. For more about worksheets, refer to
(on page 6-183).
To abort a test, select
Stop
. All test and action execution stops immediately.