Model 4200A-SCS Parameter Analyzer Reference Manual
Section 7: Keithley Configuration Utility (KCon)
4200A-901-01 Rev. C / February 2017
7-17
Set up KXCI for GPIB control
To set up GPIB control:
1. Select
KXCI Settings
. The
KXCI Settings
dialog box is displayed, as shown in the following
figure.
2. Set Communications to
GPIB
.
3. Set the
GPIB Address
. This is the primary address of the 4200A-SCS when operating under
KXCI control. If the selected GPIB address conflicts with the GPIB address of another system
component, a red exclamation-point symbol (!) is displayed next to the selected address.
4. Set the
Reading Delimiter
to determine the output data delimiter characters that are added to
the end of each KXCI output message:
Select
String Terminator
to use a character.
Select
Comma
to terminate output data with a comma (,).
5. If you selected String Terminator, select the type of
Character
:
None
to use no character.
CR
to use a carriage return.
LF
to use a line feed.
CR/LF
to use a carriage return and line feed character sequence.
6. If
String Terminator
is selected, select
EOI ON
or
EOI OFF
. The EOI setting determines if the
4200A-SCS asserts the GPIB End Or Identify (EOI) signal with the last byte of each output data
message.
7. Select
OK
.
Figure 416: KXCI Settings dialog box