Model 4200A-SCS Parameter Analyzer Reference Manual
Section 8: Keithley User Library Tool (KULT)
4200A-901-01 Rev. C / February 2017
8-5
Data type field
The data type field specifies the parameter data type. Click the arrow at the right of the data type field
to choose from a list of the following data types:
•
char:
Character data
•
char*:
Pointer to character data
•
float:
Single-precision floating point data
•
float*:
Pointer to single-precision floating point data
•
double:
Double-precision data
•
double*:
Pointer to double-precision point data
•
int:
Integer data
•
int*:
Pointer to integer data
•
long:
32-bit integer data
•
long*:
Pointer to 32-bit integer data
•
F_ARRAY_T:
Floating point array type
•
I_ARRAY_T:
Integer array type
•
D_ARRAY_T:
Double-precision array type
I/O field
The I/O field defines whether the parameter is an input or output type. Click the arrow to the right of
the I/O field to select from the input and output selections.
Default, min, and max fields
The
Default
field specifies the default value for a non-array (only) input parameter.
The
Min
field specifies the minimum recommended value for a non-array (only) input parameter.
When the user module is used in a Clarius user test module (UTM), configuration of the UTM with a
parameter value smaller than the minimum value causes Clarius to display an out-of-range message.
The
Max
field specifies the maximum recommended value for a non-array (only) input parameter.
When the user module is used in a Clarius UTM, configuration of the UTM with a parameter value
larger than the maximium value causes Clarius to display an out-of-range message.