Section 8: Keithley User Library Tool (KULT)
Model 4200A-SCS Parameter Analyzer Reference Manual
8-48
4200A-901-01 Rev. C / February 2017
Add a new UTM to the ivswitch project
To add rdsonAvg to the ivswitch project:
1. Choose
Select
.
2. Select
Projects
.
3. In the Search box, enter
ivswitch
and select
Search
. The Library displays the I-V Switch Project
(ivswitch).
4. Select
Create
. The
ivswitch
project replaces the previous project in the project tree.
5. Select the
Tests
tab.
6. For the Custom Test, select
Choose a test from the pre-programmed library (UTM)
.
7. Drag
Custom Test
to the project tree. The test has a red triangle next to it to indicate that it is not
configured.
8. Select
Rename
.
9. Enter
rdsonAvg
and press
Enter
.
10. In the project tree, drag
rdsonAvg
to the
4terminal-n-fet
device, after the
rdson
test.
11. Choose
Configure
.
12. In the Test Settings pane, from the User Libraries list, select
KI42xxulib
.
13. From the User Modules list, select
Rdson42XX
.
14. Select
Save
.
The project tree for the
ivswitch
project with
rdsonAvg
added is shown in the following figure.
Figure 454: Project tree with rdsonAvg added to 4terminal-n-fet device