Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-34
4200A-901-01 Rev. C / February 2017
YB
This command configures the Y2-axis of the graph.
Usage
YB '
AAAAAA
',
B
, ±
CCCC.CCC
, ±
DDDD.DDD
AAAAAA
The SMU channel name for the Y2-axis, up to 6 characters; must be one of the
SMU channel names that you specify on the channel definition (
DE
) page
B
Y2 axis scale type:
•
Linear scale:
1
•
Log scale:
2
•
Log scale absolute value:
3
CCCC.CCC
Y2-axis minimum value
•
Volts: ±9999
•
Amps: ±999
DDDD.DDD
Y2-axis maximum value
•
Volts: ±9999
•
Amps: ±999
Example
YB 'I2', 2, 100E-9, 1E-3
The command string:
•
Specifies that values from SMU channel I2 are to be plotted on the Y2-axis.
•
Sets up the Y2-axis to be scaled logarithmically between 100 nA and 1 mA.
Also see
Measurement control page (MD)
To send the command string to control measurements, select the measurement control page by
sending the command:
MD