Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-37
SV
This command saves a program file or data file.
Usage
SV '
A
BBBBBB
'
SV '
A
BBBBBB
CCCCCCCC
'
A
File type:
•
Program file:
P
•
Data/Program file:
D
BBBBB
Name of file (up to 6 characters)
CCCCCCCC
Comment (up to 8 characters)
Details
You must separate the file name and the comment from each other by a space.
When saving a program file, the present instrument settings are stored in a file at the directory path:
C:\s4200\sys\KXCI
You specify the name for the file.
The
get
command string is used to acquire the saved file.
Example
SV 'P Setup1'
This command string saves the command sequence as a program file named
Setup1
.
Also see
(on page 9-37)
GT
This command acquires (loads) the saved data file or program file.
Usage
GT '
A
BBBBBB
'
A
File type:
•
Program file:
P
•
Data/program file:
D
BBBBB
Name of file (up to 6 characters)
Details
The file type and file name must be separated by a space.
For a program file, this command launches the program. For a data file, it opens the files. When the
saved program file is recalled, the instrument returns to the settings stored in that file.
The save command string is used to save instrument settings or to store data acquired in a test.