Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-48
4200A-901-01 Rev. C / February 2017
EC
This command sets the condition to exit the test if compliance is reached.
Usage
EC
A
A
Action on compliance:
•
Off (do not exit if compliance is reached):
0
•
On (exit if compliance is reached):
1
Example
EC 1
This command enables exit on compliance.
Also see
None
EM
This command switches between 4145 Emulation and 4200A command sets.
Usage
EM
A
,
B
A
Mode:
•
4145 Emulation:
0
•
4200A:
1
B
Which sessions:
•
This session only:
0
•
This and all subsequent sessions (writes to KCon file):
1
Example
EM 0,1
This selects the 4145 Emulation command set permanently.
Also see