Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-53
Data format for user mode readings
For user mode operation, use the
TI
or
TV
command string to trigger and make a reading. After
sending the
TI
or
TV
command string and addressing the 4200A-SCS to talk, the output data string is
sent to the computer in the following format:
XYZ
±
N
.
NNNN
E±
NN
Data status
The hierarchy for data status is
L
,
V
,
X
,
C
,
T
,
N
, where
X
is:
•
N
: Normal
•
L
: Interval too short
•
V
: Overflow reading (A/D converter saturated)
•
X
: Oscillation
•
C
: This channel in compliance
•
T
: Other channel in compliance
Measure channel
Voltage measure mode specified (
Z
=
V
; see below)
Y =
A
SMU1
B
SMU2
C
SMU3
D
SMU4
E
VM1
F
VM2
G
SMU5
H
SMU6
I
SMU7
J
SMU8
K
VM3
L
VM4
M
VM5
N
VM6
O
VM7
P
VM8
Current measure mode specified (
Z
=
I
; see below)
Y =
A
SMU1
B
SMU2
C
SMU3
D
SMU4
E
SMU5
F
SMU6
G
SMU7
H
SMU8