Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-70
4200A-901-01 Rev. C / February 2017
:CVU:SWEEP:LISTDCV
This command configures the selected CVU card to sweep arbitrary DC voltage points and sample Z-
measurements.
Usage
:CVU:SWEEP:LISTDCV
pt1
,
pt2
,
pt3 ... ptn
pt1
,
pt2
,
pt3
…
ptn
Sweep points (1 to 4096)
Details
Only starting voltage sweep points are set with this command.
The other parameters (AC voltage level, frequency, and so on) are set by their respective commands.
Valid voltage levels are -30 V to +30 V.
Also see
None
:CVU:TEST:RUN
This command starts a CVU test on the specified card.
Usage
:CVU:TEST:RUN
Details
Use the serial poll byte to determine when not busy or data ready.
Also see
None
:CVU:TEST:ABORT
This command stops a running KXCI CVU test.
Usage
:CVU:TEST:ABORT
Details
This command terminates all ongoing processes and returns the 4200A-SCS to the idle state. Data
that results from the ongoing processes may be corrupt. Note that this command is not appropriate to
stop a CVU KXCI test running from the Remote UTM mode (for more information, refer to
(on page 9-81)).
This command is not valid in User Mode.
Also see
None