Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-74
4200A-901-01 Rev. C / February 2017
:CVU:SPEED
This command sets the measurement speed for the selected CVU card.
Usage
:CVU:SPEED
speed
:CVU:SPEED 3,
delay_factor
,
filter_factor
,
aperture
speed
Applies one of four defined speed selections:
•
0: Fast
•
1: Normal
•
2: Quiet
•
3: Custom
delay_factor
0 to 100
filter_factor
0 to 100
aperture
0.006 to 10.002 PLCs
Details
The parameters
delay_factor
,
filter_factor
, and
aperture
are only used if
speed
is set to
custom.
Also see
None
:CVU:ACV
This command sets the AC drive level for specified CVU card.
Usage
:CVU:ACV
aclevel
aclevel
10 mV to 100 mV
Details
If in user mode, this takes immediate effect.
If in system mode, the value is buffered for use in all sweeps except AC Voltage sweep (which is set
with the
:CVU:SWEEP:ACV
command).
Also see
(on page 9-68)