Section 13: LPT library function reference
Model 4200A-SCS Parameter Analyzer Reference Manual
13-68
4200A-901-01 Rev. C / February 2017
devint
is implicitly called by
execut
and
tstdsl
.
devclr
is implicitly called by
clrcon
.
The SMUs return to the following states:
•
100 µA and 10 V ranges
•
Autorange on
•
Voltage source
•
0 V DCV bias
The 4210-CVU returns to the following states:
•
30 mV
RMS
AC signal
•
0 V DCV bias
•
100 kHz frequency
•
Autorange on
•
Cable length compensation set to 0 m
•
Open/Short/Load compensation disabled
The 4225-PMU or 4220-PGU returns to the following states:
•
The pulse mode is maintained. For example, if the pulse card is in Segment Arb mode, it will still
be in Segment Arb mode after the
devint
process is complete.
•
5 V and 10 mA ranges
•
If in pulse mode:
Period of 1.0
µ
s
Transition Times (Rise and Fall) of 100
n
s
Width of 500
n
s
Voltage high and low of 0 V
Load of 50 Ω
•
If in segmented arb mode, Start Voltage is 0 V
•
If in arbitrary waveform mode, Table Length is 100
Also see
(on page 13-199)
(on page 13-9)
(on page 13-11)
(on page 13-92)
(on page 13-67)
(on page 13-24)