Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Section 2: Connections
4200A-PMU-900-01 Rev. A December 2020
2-19
Offset current compensation
Error currents can be introduced into your pulsed measurements setup by PMUs. PMU offset
compensation reduces error currents by subtracting measurements taken at 0 V from all subsequent
readings.
Perform connection compensation
To compensate for connections:
1. In Clarius, select
Tools
. The Clarius Tools dialog box opens.
Figure 19: Clarius Tools dialog box
2. Select
PMU Connection Compensation
. The Short and Offset Current Connection
Compensation Values and Defaults dialog box opens.
3. From the PMU list, select the PMU that you want to perform compensation for.