Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Section 3: Setting up PMUs and PGUs in Clarius
4200A-PMU-900-01 Rev. A December 2020
3-29
Figure 58: Two-channel test (stepping and sweeping) with three points (only PMU1-1
displayed)
Configuring the preview to all channels displays both the sweeper and stepper for the three-pulse test
(see following figure).
The Number of Pulses parameter determines the number of pulses that are output and measured for
each attempted point in the sweep or step. The PMU amplitude sweep and step example
(two-channel) test figure shows PMU1-1 (the stepper) with the number of pulses changed to three.
Notice that each sweep point of the displayed waveform in the figure has three pulses (number of
pulses set to three). The light blue waveform in the figure has one pulse (number of pulses set to one).
Because the cursor in the lower graph always contains the complete waveform of each point in a test,
the top graph shows the three pulses. The following figure shows both channels for the sweep and
step test with the number of pulses set to three. Note that the value for the number of pulses does not
affect the number of sweep or step points in the test.