Section 3: Setting up PMUs and PGUs in Clarius
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
3-32
4200A-PMU-900-01 Rev. A December 2020
Figure 61: Zooming (Expanded View graph)
Scrolling the magnified area
To move the viewable area of the graph after you zoom in on a graph, hold down Shift while dragging
the mouse.
Entire Test zoom
The zoom feature is also supported on the lower Entire Test waveform graph in the same manner as
in the Expanded View. After you zoom in, you can also scroll (or move) the viewable area of the
graph. See the following figure for a view of moving the lower graph using the mouse pointer. Note
the gaps between the pulse waveforms shown in the
(on page 3-30) figure;
gaps also exist in the following figure. These gaps indicate the time between sweep points where the
PMU performs calculations for the test while the pulse channels output 0 V. To learn more about how
these gaps relate to how the PMU handles DUT load variation and measure ranging during a test,
How LLEC adjusts pulse output to the target levels
(on page 2-22).