Model 4200A-SCS Parameter Analyzer Reference Manual
Appendix K: Using a Signatone CM500 Prober
4200A-901-01 Rev. C / February 2017
K-13
3. Select
Probe Station
.
4. Select
OK
. The KCon navigation displays PRBR1 in the Configuration Navigator and the
properties.
Figure 820: Use KCon to select a prober
5. For the Model, select the
Signatone CM500 (WL250) Prober
.
6. Make sure the
Number of Pins / Positioners
is correct. The number of pins defined here
determines the pins that are available to assign to a switch matrix card column.
7. Save the configuration.
8. Exit KCon.
Clarius project example
To set up a new prober project:
1. Start
Clarius
.
2. Choose
Select
.
3. Select the
Projects
tab.
4. Drag
New Project
into the project tree.
5. Choose
Yes
to create a new project.
6. Rename the project
PRB_CM500
.
7. Select the
Wafer Plan
tab.
8. Drag
Site
to the project tree.
9. Select the
Actions
tab.
10. In the Search box, enter
prober
.
11. Drag the
Prober Initialization (prober-init)
action to the project tree. Make sure it is under the
subsite.