Appendix K: Using a Signatone CM500 Prober
Model 4200A-SCS Parameter Analyzer Reference Manual
K-16
4200A-901-01 Rev. C / February 2017
Create a test in the subsite level:
1. Choose
Select
.
2. Choose the
Tests
library.
3. Select a test for the device on your wafer.
4. Add the test to the subsite. When you add a test, an appropriate device is automatically added.
You can also add a device from the Devices library and then add a test to the device.
5. Choose the
Actions
library.
6. Add the
Prober Move to Next Site (prober-move)
action.
7. Drag
prober-contact
so that it is immediately before your test.
8. Drag
prober-move
so that it is immediately after your test.
9. Select
Configure
.
10. For
prober-move
, set the
inknumber
to 1 if you need to trigger inker 1; otherwise, set it to 0.
Figure 823: prober-next in the project tree
Probesites Clarius project example
On the 4200A-SCS, use Clarius to open and run the
probesites
project using the new
configuration file, which allows you to execute the project for this prober. This project uses a Series
700 Switching System and the
connect
action to change the instruments connected to each pin
without changing the physical configuration.
To set up
probesites
:
1. In Clarius, choose
Select
.
2. Select the
Projects
library.
3. In the Search box, enter
probesites
.
4. Create the
probesites
project.
5. Select
Configure
.
6. Set the
prober-init
mode to
6
.
7. Set the
subprobtype
. If the CM500 prober is presently not at its first site, set the
subprobtype
to
1
; otherwise, set it to
0
.