Section 5: Pulse measure and pulse generator units
Model 4200A-SCS Parameter Analyzer Reference Manual
5-34
4200A-901-01 Rev. C / February 2017
Load-line effect compensation (LLEC) for the PMU
Load-line effect compensation (LLEC) is only performed for standard pulse I-V testing using PMU
measure ranges. It is not performed when using 4225-RPM measure ranges (
≤
1 mA). The active
RPM circuitry provides its own analog LLEC (assuming there is a short cable from the RPM to the
DUT).
The basic pulse output system is a series circuit that consists of the pulse generator resistance (fixed
at 50
Ω), interconnect (cabling and pin-to-pad) resistance, and the resistance of the DUT. In this
series circuit, the sum of the voltage drops across these components is equal to the output voltage of
the pulse generator. Therefore, if the resistance of the DUT changes, the voltage seen at the DUT
also changes. This effect is called the load-line effect. See
DUT resistance determines pulse voltage
(on page 5-61) for details on how the resistance of the DUT affects the voltage across it.
For example, consider a PMU set to outp
ut voltage to a 50 Ω load (DUT). For this default setting, the
pulse card outputs twice the programmed pulse voltage. If the interconnect resistance is negligible (0
Ω), half the pulse card voltage appears across the internal pulse card resistance (50 Ω) and the other
half (which is the programmed pulse voltage) appears across the 50 Ω DUT. For example, if the pulse
card is programmed to output a 5 V pulse, the pulse card sources a 10 V pulse. Five volts will drop
across the internal 50 Ω pulse card resistance and 5 V will appear at the 50 Ω DUT.
Methods to compensate for load-line effect
The methods to compensate for load-line effect include:
•
Use the built-in load-line effect compensation (LLEC) in the PMU for the standard 2-level pulse
mode. Ideally (when LLEC is enabled), the PMU adjusts its output levels such that the
programmed output voltage appears at the DUT. For ITMs, see
(on page 5-34). For UTMs, use the
(on page 13-108) function to control LLEC.
•
Manual adjustment of the PMU output until the target pulse level is measured across the DUT.
For a pulse sweep, this manual process must be repeated for every step in the sweep.
If you are not using LLEC, you can manually set the output impedance to match the impedance of the
DUT. For example, if the impedance of the DUT is 1
kΩ, set the output impedance to 1 kΩ. Maximum
power transfer is achieved when the DUT impedance matches the output impedance setting. In a real
test environment, you may not know the exact resistance value of the DUT, and you will have the
added affect of cabling and pin-to-pad resistance (typically 3
Ω to 10 Ω). The output impedance can
be set from an ITM (see
Disable LLEC and set the output impedance
or from a UTM (see
(on page 13-145) function).