Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-169
Stress measurements
You can do stress measurements for each device.
If the device is being stressed by DC voltage, the DC current is measured.
If the device is being stressed by DC current, the DC voltages is measured.
Stress measurements are placed in the subsite Analyze sheet in the Stress tab.
When a device is being stressed by AC voltage (VPU), the current or voltage cannot be measured by
the VPU.
You can select options to control the stress measurements. The options are:
•
Do Not Measure: Do not make the specified measurement.
•
First Stress Only: Make the specified measurement on the first stress cycle only.
•
Every Stress Cycle: Take the specified measurement on every stress cycle.
Clear, copy, and paste to all sites
Clear
: Click
Clear
to clear all stress properties data for the displayed device. It:
•
Sets all voltage and current values to
0
•
Sets device pin number assignments to
0
•
Sets Stress Measurements to Do Not Measure
•
Disables all Targets (clears Target Values)
Copy
and
Paste
: Copy and Paste allow properties settings for one device to be copied and pasted
into the properties dialog box for a different device. It can also be used to copy and paste settings into
a different site.
To use copy and paste:
1. On the Subsite Stress Properties dialog box, select
Copy
.
2. If pasting to a different site, select the site.
3. Use the
Next Device
or
Prev Device
button to display the properties dialog box for the device to
be changed.
4. Select
Paste
to overwrite the existing properties with the new properties.
Paste to All Sites
: After copying the properties for the Subsite Stress Properties dialog box, select
Paste to All Sites
to overwrite the properties for all available sites.
You can also use the Copy and Paste options in the project tree to copy a subsite structure,
including the Stress Properties settings.