Status codes
The 16 basic codes used for 4210-CVU measurement status are listed in table below. Each code is
represented as a 32-bit hexadecimal value (0x).
CVU measurement status codes (CVU1S)
#
Code
Description
1
000000
mr
No faults
2
8xxx00
mr
Measurement timeout occurred
3
01xx00
mr
CVH1 current measurement overflow
4
02xx00
mr
CVH1 voltage measurement overflow
5
03xx00
mr
CVH1 I and V measurement overflow
6
08xx00
mr
CVH1 ABB not locked
7
09xx00
mr
CVH1 ABB not locked, I measurement overflow
8
0Axx00
mr
CVH1 ABB not locked, V measurement overflow
9
0Bxx00
mr
CVH1 ABB not locked, I and V measurement overflow
10
xx0100
mr
CVL1 I measurement overflow
11
xx0200
mr
CVL1 V measurement overflow
12
xx0300
mr
CVL1 I and V measurement overflow
13
xx0800
mr
CVL1 ABB not locked
14
xx0900
mr
CVL1 ABB not locked, I measurement overflow
15
xx0A00
mr
CVL1 ABB not locked, V measurement overflow
16
xx0B00
mr
CVL1 ABB not locked, I and V measurement overflow
As shown in this table, the
mr
value is the last two digits of each code.
mr
value
00
Lowest range (1 µA) used for the impedance measurement
01
Middle range (30 µA) used for the impedance measurement
02
Highest range (1 mA) used for the impedance measurement
Measurement status notes
Whenever a fault occurs, run the Confidence Check utility before performing any other
troubleshooting actions (see
(on page 4-19) for details).
Measurement timeout
: Indicates that the measurement was not received after a set time (total
aperture). This timeout error may indicate that there is an issue with the 4210-CVU card. Try resetting
the hardware and running the project test again. If this error reoccurs, contact Keithley Instruments.
To reset the hardware:
1. Select
Start
.
2. Type
resethw
.
3. Select the instruments that need reset.
4. Click
Reset.
I measurement overflow
: Try a higher I measure range (or Auto) or try a lower AC drive voltage.
V measurement overflow
: Try a lower DC bias voltage.
ABB not locked
: Auto Balance Bridge was not locked when the measurement was made. The
readings and calculation results may not be accurate.