Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-254
4200A-901-01 Rev. C / February 2017
LOGFIT Formulator function
Performs a base-10 log-linear fit.
Usage
LOGFIT(
VX
,
VY
,
STARTPOS
,
ENDPOS
)
VX
The name of any column (vector) listed under Columns
VY
The name of any column (vector) listed under Columns
STARTPOS
For the range of
X
and
Y
values to be logarithmically fitted, the row number (index)
of the starting values
ENDPOS
For the range of
X
and
Y
values to be logarithmically fitted, the row number (index)
of the ending values
Details
Performs a base-10 log-linear fit as follows:
•
Fits the following logarithmic relationship to a specified range of values in two columns (vectors)
(one column,
VX
, containing
X
values and the other column,
VY
, containing
Y
values):
Y = L LOGFITB * log(X)
where
LOGFITA
and
LOGFITB
are fit constants.
•
Using the above logarithmic relationship, returns a new column (vector) containing
Y
values
calculated from all
X
values in column
VX
.
If a
VX
or
VY
value at either
STARTPOS
or
ENDPOS
is an invalid number (that is, the value is
#REF
),
the function will not return a valid result.
Example
GOODFIT = LOGFIT(GATEV, DRAINI, 30, 50)