Model 4200A-SCS Parameter Analyzer Reference Manual
Section 6: Clarius
4200A-901-01 Rev. C / February 2017
6-307
cvucompulib user library
The
cvucompulib
user library contains user modules for collecting 4210-CVU compensation data.
The next table lists and briefly describes the user modules.
If your configuration includes a 4200A-CVIV, use the
cvivulib
user library instead of this one.
cvucompulib user modules
User module
Description
cvu_ConstantsFileSelect
Selects the constants file that is used for a CVU ITM or UTM. The
file must be created using
cvu_OSLcomp_collect
.
cvu_OSLcomp_collect
Collects the open, short, and load compensations of the CVU
instrument as selected. It generates a file that contains the open,
short, and load compensation values to apply to the CVU readings
that are returned from an ITM or UTM.
DLCP user library
The
DLCP
user library contains a user module for making C-V measurement for drive-level
capacitance profiling (DLCP). The next table lists and briefly describes the user module.
DLCP user module
User module
Description
ACSweep
This module allows you to make C-V measurements for drive-level
capacitance profiling (DLCP) using the 4210-CVU. During this measurement,
the applied AC voltage is sweeping while the capacitance is measured. The
total applied voltage (AC and DC) is kept constant. The total applied voltage is
defined as the DC voltage minus 1/2 the p-p AC voltage.