Section 7: Keithley Configuration Utility (KCon)
Model 4200A-SCS Parameter Analyzer Reference Manual
7-10
4200A-901-01 Rev. C / February 2017
Update preamplifier, RPM, and CVIV Configurations
After adding or removing a preamplifier, RPM, or 4200A-CVIV, you must update the system by
selecting the
Update Preamp, RPM, and CVIV Configuration
option.
Update the preamplifier configuration
For more information on adding or removing preamplifiers, refer to
To update the preamplifier configuration:
1. Remove or reconnect a preamplifier to the SMU.
2. Select the
Update Preamp, RPM, and CVIV Configuration
option.
3. Select
Save
.
Update the RPM configuration
The KCon feature Update Preamp, RPM, and CVIV Configuration must be run:
•
Whenever an RPM is connected or disconnected from the 4200A-SCS
•
Whenever RPM input connections for a SMU, CVU, or PMU are changed
When using interactive test modules (ITMs) for DUT testing, you can configure RPM switching in
KCon. After you connect the PMU, SMUs, and CVU to the RPM, update the RPM configuration in
KCon. When you then open Clarius, the system will detect the instrument connections to the RPM.
When an ITM is run, Clarius will automatically close the appropriate switches to connect the PMU,
SMU, or CVU to the output of the RPM. Use the following procedure to configure RPM switching:
For more information on adding or removing RPM connections, refer to
(on page 5-
To update RPM configuration:
1. Turn off system power.
2. Disconnect line power.
3. Connect the correct RPM to the correct PMU channel.
4. If the system contains SMUs: Connect any SMU to any RPM using either one or two triaxial
cables. Repeat for all RPMs in the system.
5. If the system contains a CVU: Connect CVU High to one RPM, and CVU Low to the other RPM.
6. Turn on the system power.
7. Start KCon.
8. Select the
Update Preamp, RPM, and CVIV Configuration
option.
9. Select
Save
.
10. Exit from KCon.
If you are using an interactive test module (ITM) for testing, the RPM automatically selects the PMU,
SMU, or CVU. If you are using a user test module (UTM), you must select the correct RPM settings.
See