Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-16
4200A-901-01 Rev. C / February 2017
VM
This command defines channels that are used as voltmeters.
Usage
VM
A
, '
BBBBBB
'
A
Voltmeter channel: 1 to 9; the assigned value for a voltmeter depends on how
instruments are mapped in KCon
BBBBBB
User-specified name (up to 6 characters)
Details
KXCI allows up to eight source-measure units (SMUs) to function solely as voltmeters. You can use
any channel for any voltmeter function between VM1 and VM9. For example, in a system containing
four SMUs, you can use
SMU3
as
VM7
.
If you do not define one of the 4200A-SCS SMUs to emulate a VM, attempts to measure voltages
through the nonexistent VM result in data values of
9.000e+37
.
If nothing is specified after the prefix and channel number, the channel is not used.
To measure voltage using the 4145B
VM1...VM
n
function, define one of the 4200A-SCS SMUs to
emulate
VM
.
Example
VM1, 'VM1'
This command string defines the channel used for VM1 (specifies the name
VM1
).
Also see
(on page 7-16)
Source setup page (SS)
Use the command strings for the SS page for the following operations:
•
VAR1 setup
•
VAR2 setup
•
VAR1' setup
•
List sweep setup
•
Constant SMUs setup
•
Constant VS setup
•
Set sweep hold time
•
Set sweep delay time
In order to send the following command strings to the 4200A-SCS, you must first select the source
setup page by sending the command:
SS