VP and IP
This command sets up the VAR2 step sweep.
Usage
AA
±
BBB.BBBB
, ±
CCC.CCCC
,
DD
, ±
EEE.EEEE
AA
±
BBB.BBBB
, ±
CCC.CCCC
,
DD
, ±
EEE.EEEE
,
FF
AA
The source mode:
•
VP
: Voltage source (
SMU
or
VS1...VS8
)
•
IP
: Current source (
SMU
only)
BBB.BBBB
Start value:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
CCC.CCCC
Step value:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
DD
Number of steps: 1 to 32
EEE.EEEE
Compliance value (see
Details
):
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
FF
VAR2 source stepper index: 1 to 4 (see
Details
)