Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-30
4200A-901-01 Rev. C / February 2017
NR
This command sets the number of readings that can be made for time domain measurements.
Usage
NR
AAAA
AAAA
Number of measurements to make:
•
4200A command set: 1 to 4096
•
4145 Emulation command set: 1 to 1024
Details
For time domain measurements, you can set the number of sample measurements that can be made.
The readings are stored in the buffer.
Example
NR 200
This command string sets up the 4200A-SCS to make 200 sample measurements.
Also see
None
DM
This command selects the Keysight 4145B display mode.
Usage
DM
A
A
Display modes:
•
Graphics display mode:
1
•
List display mode:
2
Details
The 4200A-SCS supports the Keysight 4145B graphics display mode and the Keysight 4145B list
display mode command.
The 4200A-SCS does not accept the matrix mode and schmoo mode commands (DM3 and DM4).
Example
DM1
This command string prepares the 4200A-SCS to receive graphics commands.
Also see
None