Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-41
Details
For every channel that is configured as a SMU, you must select the source mode (voltage or current)
and source output range, and set the output and compliance values.
With the voltage source mode (
DV
) specified in the command string, the output value is in volts. For
the current source mode (
DI
), the output value is in amps.
With the voltage source mode (
DV
) specified in the command string, you are setting the current
compliance. For the current source mode (
DI
), you are setting the voltage compliance. When
sourcing voltage, note that if you specify a compliance current that is below the minimum-allowable
value (1 pA with a preamplifier installed and 100 nA without a preamplifier), KXCI sets it to the
minimum allowable value.
Example
DV1, 1, 10, 10E-3
This command string configures SMU1 to source 10 V on the 20 V source range and sets current
compliance to 10 mA.
Also see
None
DS
This command specifies the channel number and the voltage output value for each voltage source.
Usage
DS
A
, ±
BBB.BBBB
A
n
, for voltage source VS
n
; range 1 to 9
BBB.BBBB
Output voltage value:
−
200.00 to +200.00
Details
KXCI allows up to nine source-measure units to function solely as voltage sources. You can use any
channel for any voltage-source function between
VS1
and
VS9
. For example, in a system containing
four SMUs, you can use SMU2 as
VS5
.
The assigned
n
value for a voltage source (
VS
n
) or voltmeter (
VM
n
) depends on how instruments are
mapped in KCon.
Example
DS1, 20
This command string sets
VS1
to output 20 V.
Also see
(on page 7-1)