Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-49
*OPT?
This command returns a result string that contains the 4200A-SCS slot configuration for KXCI.
Usage
*OPT?
Details
When the 4200A-SCS receives this command, it returns the following configuration string:
xxx
,
xxx
,
xxx
,
xxx
,
xxx
,
xxx
,
xxx
,
xxx
This string contains eight sets of characters, each represented by
xxx
. Each character set represents
the configuration of a slot in the 4200A-SCS. If the corresponding slot contains a channel,
xxx
is one
of the following:
SMU
n
,
HPSMU
n
,
SMUPA
n
,
HPSMUPA
n
,
VS
n
, or
VM
n
, where
n
is the channel number (1
to 9) and:
SMU
= Medium-power SMU without a preamplifier
HPSMU
= High-power SMU without a preamplifier
SMUPA
= Medium-power SMU with a preamplifier
HPSMUPA
= High-power SMU with a preamplifier
VS
= Voltage source only (20 V is presently supported).
VM
= Voltage meter
If the corresponding slot does not contain a channel, then
xxx
=
""
.
Also see
Ethernet command reference
The Ethernet command set includes all the GPIB commands documented in the
GPIB command
reference
(on page 9-13), plus the
SP
command. The
SP
command allows you to acquire the GPIB
spoll byte when ethernet communications is selected.
The
SP
command is not available for GPIB communications.
4200A command set only commands
The
OPT?
command string returns a result string that indicates the 4200A-SCS slot configuration for
KXCI.