Model 4200A-SCS Parameter Analyzer Reference Manual
Section 13: LPT library function reference
4200A-901-01 Rev. C / February 2017
13-15
The 4225-PMU or 4220-PGU returns to the following states:
•
The pulse mode is maintained. For example, if the pulse card is in Segment Arb mode, it will still
be in Segment Arb mode after the
devint
process is complete.
•
5 V and 10 mA ranges
•
If in pulse mode:
Period of 1
µ
s
Transition Times (Rise and Fall) of 100
n
s
Width of 500
n
s
Voltage high and low of 0 V
Load of 50 Ω
•
If in segmented arb mode, Start Voltage is 0 V
•
If in arbitrary waveform mode, Table Length is 100
Also see
(on page 13-199)
(on page 13-9)
(on page 13-11)
(on page 13-92)
(on page 13-67)
(on page 13-24)
disable
This command stops the timer and sets the time value to zero (0).
Usage
int disable(int
instr_id
);
instr_id
The instrument identification code of the timer module (
TIMER
n
)
Details
Timer reading is also stopped.
Sending
disable(TIMER
n
)
stops the timer and resets the time value to zero (0).
Also see
(on page 13-16)