Model 4200A-SCS Parameter Analyzer Reference Manual
Section 13: LPT library function reference
4200A-901-01 Rev. C / February 2017
13-133
devclr
This command sets all sources to a zero state.
Usage
int devclr(void);
Details
This command clears all sources sequentially in the reverse order from which they were originally
forced. Before clearing all Keithley supported instruments, GPIB-based instruments are cleared by
sending all strings defined with the
kibdefclr
command.
devclr
is implicitly called by
clrcon
,
devint
,
execut
, and
tstdsl
.
For C-V testing, this command turns off the DC bias voltage.
Also see
(on page 13-199)
(on page 13-13)
(on page 13-16)
(on page 13-22)
(on page 13-50)
devint
This command resets all active instruments in the system to their default states.
Usage
int devint(void);
Details
Resets all active instruments in the system to their default states. It clears the system by opening all
relays and disconnecting the pathways. Meters and sources are reset to their default states. Refer to
the hardware manuals for the instruments in your system for listings of available ranges and the
default conditions and ranges.
The
devint
command is implicitly called by the
execut
command.