Section 5: KPulse (for Keithley Pulse Cards)
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
5-10
4200A-PMU-900-01 Rev. A December 2020
3. Select the
Waveform Type
to be created.
4. Configure the
Settings
for the selected waveform type.
5. Select
Preview
to update the preview of the waveform.
6. In the
Waveform Name
field, type in a name for the waveform. You cannot use a name that is
already used in the Scratch Pad.
7. Select
OK
to create the waveform. The new waveform is added to the Scratch Pad.
8. Repeat these steps to create another waveform in the Scratch Pad.
Custom Arb file operation: Copy waveforms into Sequencer
To copy waveforms into the Sequencer:
1. Select the
Arb Generator
tab.
2. Select
Scratch Pad
or
Sequencer
:
▪
Scratch Pad previews the waveform that is selected in the Scratch Pad.
▪
Sequencer previews enabled waveform sequences. To preview the waveforms in the
Sequencer, select
Enable Channel 1
or
Enable Channel 2
.
3. Select the scale for the graph,
Points Scale
or
Time Scale
.
4. In the Scratch Pad, select a waveform to be copied into the Sequencer.
5. Select
Channel 1
to copy the selected waveform onto the Sequencer for Channel 1 or select
Channel 2
to copy the waveform into the Sequencer for Channel 2. You can copy the same
waveform to both channels.