Section 3: Setting up PMUs and PGUs in Clarius
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
3-8
4200A-PMU-900-01 Rev. A December 2020
Force Range (PMU)
The range that is used when sourcing. Select one of the listed ranges. The source remains on the
range that is set. If you are sweeping and a sweep point exceeds the source range capability, the
source outputs the maximum level for that range. This range must be equal to or greater than the
largest value in the sweep.
Note that the 40 V range has a higher maximum output voltage and current, but does not have the
faster transition times of the 10 V range or the lower current measure ranges of the 4225-RPM.
Disable outputs at completion
Select this option to disable the outputs when the test completes.
Clear this option to leave the outputs enabled when the test completes.
Current Spot Mean High
Makes current measurements on the amplitude. Available when the Test Mode is set to Pulse I-V.
The following figure shows an example of a spot mean measurement on pulse high (amplitude) and
pulse low (base level). For a more detailed example, refer to
(on page 3-16).
Figure 35: Spot mean measurements
Current Spot Mean Low
Makes current measurements on the base level. Available when the Test Mode is set to Pulse IV.
The following figure shows an example of a spot mean measurement on pulse high (amplitude) and
pulse low (base level). For a more detailed example, refer to
(on page 3-16).
Figure 36: Spot mean measurements