Power and environmental specifi cations
For indoor use only.
Power Supply
100 VAC to 240 VAC, 50 Hz to 60 Hz
Maximum VA
1000 VA
Operating altitude
Maximum 2000 m (6562 feet) above
sea level
Operating
temperature
10 °C to 40 °C (50 °F to 104 °F),
5 % to 80 % relative humidity,
non-condensing
Storage temperature
-15 °C to 60 °C (5 °F to 140 °F),
5 % to 90 % relative humidity,
non-condensing
Carefully consider and confi gure the appropriate
output-off state, source levels, and compliance
levels before connecting the instrument to
a device that can deliver energy. Failure to
consider the output-off state, source levels, and
compliance levels may result in damage to the
instrument or to the device under test.