Model 4210-MMPC-C

Multi-Measurement Prober Cable Kit 

Quick Start Guide

Installation guidelines

 

Figure 7 shows the prober cable assemblies of four Model 4200-MMPC-C kits installed in a Cascade Microtech 
12000 series prober. 

Figure 7: Four prober cable assemblies installed in a Cascade prober 

 

To install the Cascade prober kit:  

1.  Referring to Figure 7, mount each grounding bracket onto the tube of a manipulator.  
2.  Firmly tighten the set screws to secure the brackets to the tubes.  
3.  Black jumpers are used to connect the commons of the cable assemblies together. The jumpers plug into 

the grounding brackets. Referring to Figure 7, connect the jumpers from one grounding bracket to another. 
For the basic 2-pin setup in Figure 6, it is good practice to install the second jumper. It not only improves the 
connection, but it is a good place to store the extra jumper. 

NOTE 

Figure 7 shows an example of good routing for the jumper wires. When routing the 
wires, consider the possible movements of the manipulators.  

4.  Referring to Figure 6, connect the prober manipulators to the prober bulkhead connectors using the supplied 

triax cables. Each cable is terminated with a full-triax connector on one end and a mini-triax connector on 
the other end. Figure 7 shows the mini-triax connectors mounted on the manipulators.  

Summary of Contents for 4210-MMPC-C

Page 1: ...V A package to perform pulse I V measurements This kit has been assembled for use with the Cascade Microtech 12000 prober series manipulator type DCM 200 series For C V testing this Quick Start Guide...

Page 2: ...is a one page packing list for the Model 4210 MMPC C Multi Measurement Prober Cable Kit for the Cascade MicroTech 12000 series prober A photo shows all the components of the cable kit Section 16 of th...

Page 3: ...Figure 2 shows SMUs that do not use preamps These SMUs use the supplied triax cables that are terminated with a miniature triax connector on one end and a standard triax connector on the other end If...

Page 4: ...r Cable Kit Quick Start Guide C V testing Two prober cable kits include the cable adapters needed to connect the subminiature version A SMA cables of the Model 4210 CVU to the female triax connectors...

Page 5: ...ctors to connect the Model 4200 to the bulkhead of the prober see Figures 4 and 5 Figure 4 shows the Model 4200 connections to the prober and the trigger connection between the scope card Model 4200 S...

Page 6: ...ables and adapters used for instrumentation connections Installing the prober cable kits Basic cable setup Figure 6 shows the basic cable setup for the 2 pin I V and C V testing scenarios described in...

Page 7: ...jumpers are used to connect the commons of the cable assemblies together The jumpers plug into the grounding brackets Referring to Figure 7 connect the jumpers from one grounding bracket to another Fo...

Page 8: ...Us connected to the prober bulkhead use the fundamental setup shown in Figure 8 to perform 2 pin I V testing Figure 2 shows the SMU connections to the outside of the bulkhead Figure 8 Test setup for 2...

Page 9: ...tart Guide Figure 9 Test setup for 4 pin I V testing C V testing 2 pin C V setup NOTE The setup for 2 pin C V testing requires two prober cable kits With a Model 4210 CVU connected to the prober bulkh...

Page 10: ...test setup using the Keithley Model 4200 PIV A NOTE The setup for 4 pin pulse I V testing requires four prober cable kits The Model 4200 PIV A is a factory installed package that includes a scope car...

Page 11: ...k Start Guide NOTE Details on using the Model 4200 PIV A package are provided in the Model 4200 SCS User s Manual see How to use the Model 4200 SCS to perform a pulse I V test on my device in Section...

Page 12: ...prober cable kits The Model 4225 PMU is an Ultra Fast I V Module that is an instrument card for the Model 4200 SCS The PMU has two channels of voltage pulse source with integrated simultaneous voltag...

Page 13: ...erminal device some re cabling is required for certain C V test configurations The 4225 RPM has the appropriate connector types at the input and output of the RPM enclosure to eliminate any adapters w...

Page 14: ...Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com 14 PA 1001 Rev D September 2011 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 17 Test setup for three pin devi...

Page 15: ...Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com PA 1001 Rev D September 2011 15 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 18 Test setup for four pin devic...

Page 16: ...28775 Aurora Road Cleveland Ohio 44139 1 888 KEITHLEY www keithley com 16 PA 1001 Rev D September 2011 Model 4210 MMPC C Multi Measurement Prober Cable Kit Quick Start Guide Figure 19 Test setup for f...

Reviews: