Model 6220/6221 User’s Manual
Averaging Filter, Math, and Buffer
6-9
Return to
Buffer statistics
– While in buffer RECALL, use the
EDIT/LOCAL
key (see
). Each press of this key displays the next statistic. After the last statistic
is displayed (Std Dev), pressing
EDIT/LOCAL
will display the stored readings.
When finished, press
EXIT
to return to the normal display.
Figure 6-1
Buffer recall
For
remote programming
, the commands to recall delta readings and statistics
are listed in
.
Table 6-4
Buffer commands
Command
Description
Default
TRACe:DATA?
Request all readings in buffer.
CALCulate2:FORMat <name>
Select buffer statistic:
<name> = MEAN, SDEViation, MAXimum,
MINimum, or PKPK
MEAN
CALCulate2:STATe <b>
Enable or disable calculation:
<b> = ON or OFF
OFF
CALCulate2:IMMediate
Perform the selected calculation on buffer readings.
CALCulate2:DATA?
Read the result of the buffer statistic.
PUSH TO ENTER
+123.45678mV
00000
Isrc +1.0000mA
Time 000000.001 s
Reading
Reading Number
Current Source
Timestamp
Buffer Readings
RANGE
RANGE
or
RANGE
RANGE
or
A) Front panel edit keys to display buffer readings:
Model 6220:
Model 6221:
EDIT/
LOCAL
B) Use EDIT/LOCAL key to display buffer statistics:
Min
®
Max
®
Pk-Pk
®
Avg
®
Std Dev
®
Buffer Readings
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......