Model 6220/6221 User’s Manual
Wave Functions (6221 Only)
7-5
Return to
Duty cycle
A, the duty cycle setting is the portion of the total
cycle that the wave is high relative to the period of the waveform. For a ramp
waveform
B, the duty cycle corresponds to the fraction of the total wave
cycle that is rising. A 50% duty cycle corresponds to a symmetric triangle wave.
The left waveform in B has a 100% duty cycle, the center waveform has 0% duty
cycle, and the right waveform has 50% duty cycle. A 0
°
marker setting for ramp
waveforms corresponds to the minimum output at the start of ramp-up. For sine
and square wave, a 0
°
marker setting for square and sine waves is the zero-
crossing point.
Figure 7-2
Duty cycle
Duty
Cycle
Setting
A. Square Wave Duty Cycle
100% Duty
Cycle Setting
B. Ramp Wave Duty Cycle
0% Duty
Cycle Setting
50% Duty
Cycle Setting
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......