B-4
SCPI Tables (Abridged)
Model 6220/6221 User’s Manual
Return to
Table B-6
Source command summary
Command
Description
Default
SOUR
Current source output commands:
:CLE
Clearing the current source:
[:IMM]
Set output to zero and then turn the output off.
:CURR
Current source:
[:AMPL] <n>
Set current source output (amps): -105e-3 to 105e-3
0.0
:RANG <n>
Select a fixed source range: -105e-3 to 105e-3
100e-3
:AUTO <b>
Enable or disable source auto range.
OFF
:COMP <NRf>
Set voltage compliance (volts): 0.1 to 105
10
:FILT <b>
Enable or disable analog filter.
:STAR <n>
Set start current (amps): -105e-3 to 105e-3
0
:STOP <n>
Set stop current (amps): -105e-3 to 105e-3
100e-3
:STEP <n>
Set step current (amps): 1e-13 to 105e-3
1e-2
:CENT <n>
Set center current (amps): -105e-3 to 105e-3
0
:SPAN <n>
Set span current (amps): 2e-13 to 210e-3
100e-3
:DEL <n>
Set source delay (seconds): 1e-3 to 999999.999
1
:SWE
Source sweep operation:
:SPAC <name>
Select sweep type: LIN, LOG or LIST
LIN
:POIN <n>
Set sweep points: 1 to 65535
11
:RANG <name>
Select sweep ranging: AUTO, BEST or FIX
BEST
:COUN <NRf>
Set sweep count: 1 to 9999 or INFinity
1
:CAB <b>
Enable to abort sweep on compliance: ON or OFF
OFF
:ARM
Arm the sweep.
:ABOR
Abort sweep or delta test immediately.
:LIST
Custom sweep operation:
:CURR <NRf>
[,<NRf>, ...<NRf>]
Define list of currents (amps): -105e-3 to 105e-3
:APP <NRf>
[,<NRf>, ...<NRf>]
Add current points to existing list: -105e-3 to 105e-3
:POIN?
Query number of current list points.
:DEL <NRf>
[,<NRf>, ...<NRf>]
Define list of delay values (secs): 1e-3 to 999999.999
1
:APP <NRf>
[,<NRf>, ...<NRf>]
Add delay values to existing list.
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......