Model 6220/6221 User’s Manual
SCPI Tables (Abridged)
B-5
Return to
:COMPliance <NRf>
[,<NRf>, ...<NRf>]
Define list of compliance values (volts): 1e-3 to 105
:APP <NRf>
[,<NRf>, ...<NRf>]
Add compliance values to existing list: 1e-3 to 105
:DELta
Delta operation:
:NVPR?
Query connection to 2182/2182A: 1 = yes, 0 = no
:HIGH <NRf>
Set high source value (amps): 0 to 105e-3
1e3
:LOW <NRf>
Set low source value (amps): 0 to -105e-3
-1e3
:DEL <NRf>
Set Delta delay (seconds): 1e-3 to 9999.999
2e-3
:COUN <NRf>
Set number of cycles to run: 1 to 65536 or INFinity.
INF
:CAB <b>
Enable to abort test on compliance: ON or OFF
OFF
:CSW <b>
Enable cold switching mode: ON or OFF.
OFF
:ARM
Arm Delta.
:PDEL
Pulse Delta operation:
:NVPR?
Query connection to 2182A: 1 = yes, 0 = no
:HIGH <NRf>
Set high pulse value (amps): -105e-3 to 105e-3
1e3
:LOW <NRf>
Set low pulse value (amps): -105e-3 to 105e-3
0
:WIDT <NRf>
Set pulse width (seconds): 50e-6 to 12e-3
110e-6
:SDEL <NRf>
Set source delay (seconds): 16e-6 to 11.966e-3
16e-6
:COUN <NRf>
Set number of readings to perform: 1 to 65636 or INF
INF
:RANG <NRf>
Select fixed pulse source range: BEST or FIX
BEST
:INT <NRf>
Set interval (in PLCs) for each pulse cycle: 5 to 999999
5
:SWE <b>
Enable or disable Sweep output mode.
OFF
:LME <NRf>
Set number of low measurement per cycle: 1 or 2
2
:ARM
Arm Pulse Delta.
:DCON
Differential Conductance operation:
:NVPR?
Query connection to 2182/2182A: 1 = yes, 0 = no
:STAR <NRf>
Set start value (amps): -105e-3 to 105e-3.
-10e-6
:STEP <NRf>
Set step size (amps): 0 to 105e-3.
1e-6
:STOP <NRf>
Set stop value (amps): -105e-3 to 105e-3.
+10e-6
:DELT <NRf>
Set delta value (amps): 0 to 105e-3.
1e-6
:DEL <NRf>
Set delay (seconds): 1e-3 to 9999.999.
2e-3
:CAB <b>
Enable to abort test on compliance: ON or OFF
OFF
:ARM
Arm Differential Conductance.
Table B-6 (cont.)
Source command summary
Command
Description
Default
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......